Setting The Energy Protocol; Setting The Shock Mode - Welch Allyn AED 10 User Manual

Automatic external defibrillator
Hide thumbs Also See for AED 10:
Table of Contents

Advertisement

User Manual

Setting the Energy Protocol

The Welch Allyn AED 10 energy protocol provides a sequence of three defibrillator shocks.
The default protocol setting is 200 Joules, 300 Joules, and 360 Joules for the first,
second, and third shocks, respectively. However, the energy level for each shock is
changeable.
Energy
200
200
360
Save
Cancel
To set a new energy protocol
1.
Access the Supervisor Menu. Select Protocol to display the Protocol Menu.
2. Highlight a field (first, second, or third shock) by using the Enter soft key. Press the
Next soft key to change the value of the energy for that shock.
3. Highlight Save by pressing Enter after the desired energies have been selected.
Press Enter again to confirm the change and return to the Supervisor Menu.
4. To leave the energy protocol as it was originally, press Next when the Save box is
highlighted to select the Cancel box. Press Enter to return to the Supervisor Menu.
Selectable Shock Energies for 3-Shock Protocol
First Shock

Setting the Shock Mode

The Welch Allyn AED 10 allows to choose from 3 different shock modes. These modes
are 1 Shock CPR (goes into CPR mode after every shock), 3 Shock CPR (goes into CPR
mode after every 3 shocks), and 3 Shock Analyze (for a non-shockable heart rhythm,
continuously analyzes in the background).
Shock Mode
1 Shock CPR
Save
Cancel
To set Shock Mode
1.
Access the Supervisor Menu screen. Select Shock Mode to display the Shock
Mode Menu.
Energy
200
300
360
Save
Cancel
Second Shock
150 J
150 J
200 J
200 J
300 J
Shock Mode
3 Shock CPR
Save
Cancel
Energy
200
300
360
Save
Cancel
Third Shock
150 J
200 J
300 J
360 J
Shock Mode
3 Shock CPR
Save
Cancel
Programming
47