Advertisement

56
Maintenance

Maintenance Schedule

Caution Improper maintenance may cause the Welch Allyn AED 10 to
malfunction. Maintain the Welch Allyn AED 10 as described in this manual.

General

Prepare a customized maintenance schedule for the Welch Allyn AED 10 to reflect how
the unit will be used. The schedule should account for how often the unit is used and
where it is used. In addition, the schedule should consider how familiar the operators are
with the operation of the unit. Here are some guidelines for preparing a maintenance
schedule at your facility:
Frequency
After each use
and during
each inspection
After each use

Checklists for Preparedness

Use FDA Checklist and Automated External Defibrillators: Operator's Checklist to help
maintain the Welch Allyn AED 10 in an operation-ready condition.
An important part of a successful maintenance program is the creation of a maintenance
log in which information is recorded on a regular basis. The log provides a:
Record of the maintenance performed, who performed it, and when it was
completed.
Schedule of periodic requirements such as calibration and certification.
Tracking of accessories, such as batteries, that require periodic testing and
replacement.
Observe
Check the status indicator. Verify that the
"ready to use" status indicator on the front
of the unit is operating.
Inspect exterior of Welch Allyn AED 10 and
pad connector for any signs of damage.
Make sure that all supplies and
accessories are present and in proper
operating condition. Inspect the
defibrillation pad packages, and battery
packs for "install before dates" and any
signs of damage.
Check battery
Welch Allyn AED 10 Automatic External Defibrillator
Action
If the status indicator is flashing, replace the battery. If a
solid red symbol appears, turn on the AED 10, if possible,
and note the self-test results. If a "Self-Test Failure" is
indicated, remove AED 10 from service and contact
Welch Allyn customer service. If the AED 10 will not
power on, replace the battery as it is probably drained.
Clean the Welch Allyn AED 10. If damaged, remove AED
10 from service and contact Welch Allyn customer
service.
Do not use damaged or expired supplies or accessories.
Replace any used supplies and accessories such as
razors, gloves, and pads.
Check the battery capacity shown in the battery gauge on
the display. If "Low Battery" is indicated, remove the
battery and replace with a new battery. Dispose of the
battery properly according to local authority disposal
standards.
Always verify that a fully charged spare battery is
available for use.