Programming the AED 10

Due to software upgrades or changes in functionality, the Welch Allyn AED 10 may require
reprogramming at some time in its service life. In order to program the AED 10, the user
must enter programming mode by selecting Program from the Supervisor Menu.
Note
Programming the AED 10 should only be performed by qualified sales or service
representatives.
Program
Yes
Cancel
To program the AED 10
1.
Access the Supervisor Menu screen. Select Program to display the Program Menu.
2. Press the Enter soft key to restart the unit in program mode.
3. To leave the program as it was originally, press Next when the Yes box is highlighted
to select the Cancel box. Press the Enter soft key to return to the Supervisor Menu.
Note
After reprogramming the AED 10, the battery must be removed to turn off the
power. When the battery is reinserted, the AED 10 will perform a self-test and
then power off.

Restoring the Factory Default Settings

The Welch Allyn AED 10 has the option to revert all user-configurable settings back to the
original factory settings. Using this option completely erases all settings, including
language, which may have been set by the supervisor.
Restore Defaults
Yes
Cancel
To restore factory defaults
1.
Access the Supervisor Menu. Select Restore Defaults to display the Restore
Defaults Menu.
2. The Yes box will be highlighted. To restore defaults, press Enter.
3. To not restore defaults, press Next to highlight Cancel. Press Enter to return to the
Supervisor Menu.
Table 1. Factory default settings
Parameter
Volume
Password
CPR Timer
Restart in Program Mode
Default setting
3
123
120
Programming