Defibrillator - Welch Allyn AED 10 User Manual

Automatic external defibrillator
Hide thumbs Also See for AED 10:
Table of Contents

Advertisement

User Manual

Defibrillator

Output
Energy Sequence (user configurable)
Charge Time from "Shock Advised"
Analysis Time
Combined Analysis and Charge Time
from power up
after 6 shocks
Audible Prompts
Visual Prompts
Controls
Waveform Details
Output Energy Accuracy
MRL Orbital
Biphasic Truncated exponential
Shock 1: 150 J, 200 J
Shock 2: 150 J, 200 J, 300 J
Shock 3: 150 J, 200 J, 300 J, 360 J
4 sec. to 150 J
8 sec. to 200 J
15 sec. to 360 J
4-16 sec.
From start of analysis to shock ready:
Less than 20 seconds to 200 J
Less than 30 seconds to 360 J
18 audible prompts
20 text screen prompts
Four buttons - On/Off, Shock, and 2 software-configurable buttons
The table below provides details of the biphasic truncated exponential waveform
delivered by the AED 10 (set to Emax) when connected to resistive loads of 25
through 175 Ω. The waveforms are characterized by typical values for peak
current (I
), duration of the first output phase, and duration of the second output
p
phase. Values are within 10%.
I
p1
phase 1
phase 2
I
p2
Patient
I
p1
Impedance (Ω)
(Amps)
25
52.2
50
26.5
75
17.5
100
13.3
125
10.5
150
8.9
175
7.5
±10% or 1 J (whichever is greater) into 50 Ω impedance
±15% or 1 J (whichever is greater) into any impedance from 25 Ω to 125 Ω
Typical transthoracic impedance with properly applied Welch Allyn defibrillation
pads is under 125 Ω.
Specifications
I
t
t
p2
phase1
phase2
(Amps)
(ms)
(ms)
35.7
5.6
3.8
16.9
11.6
7.7
11.8
16.6
9.9
9.3
18.8
9.9
7.7
19.9
10.3
6.8
20.2
10.3
6.0
20.4
10.3
67