44
Programming

Accessing the Supervisor Menu

The Supervisor Menu is accessed from the Main Menu.
Main Menu
Volume
Contrast
Date
Time
Log
Supervisor
To access the Supervisor Menu
1.
While in the Main Menu, press the Next soft key to highlight Supervisor, and then
press the Enter soft key to go to the Password Menu.
2. Enter the Password by using Next to scroll to the appropriate number and use Enter
to shift to the next digit. After the third number has been selected, highlight Save by
pressing Enter.
3. Press Enter again while in the Save box. If the password is correct, the Supervisor
Menu will appear. If the password is incorrect, the Main Menu will appear.
4. To exit the Password prompt, press Next when the Save box is highlighted to select
the Cancel box. Press the Enter soft key to return to the Main Menu.
The default supervisor password is 1 2 3.
Note

Supervisor Menu Items

Supervisor (1)
CPR
Language
Protocol
Shock Mode
Program
More
Back
CPR
Language
Protocol
Shock Mode
Password
Password
000
123
Save
Save
Cancel
Cancel
Supervisor Menu
Restore Defaults
Unit ID
Dept
Change Code
Bit Rate
Utility
Back
Set the value for CPR Timer and/or CPR Tempo.
Select text and audio language. The Welch Allyn AED 10 will
reboot into automated mode to activate selected language.
Select the energy level protocol. Default protocol is 200, 300,
360 Joules.
Selects among 1 Shock CPR (default), 3 Shock CPR, or 3 Shock
Analyze mode.
Welch Allyn AED 10 Automatic External Defibrillator
Supervisor
CPR
Language
Protocol
Shock Mode
Program
More
Back