Battery Care; Care And Storage; Electrodes - Welch Allyn AED 10 User Manual

Automatic external defibrillator
Hide thumbs Also See for AED 10:
Table of Contents

Advertisement

6
Safety
Caution Do not repeatedly charge and discharge the defibrillator in rapid
succession.
Caution Improper maintenance can cause improper performance.

Battery Care

Caution Use only Welch Allyn batteries.
Caution Always verify remaining capacity of a non-rechargeable battery after
use.
Caution Replace the battery at 5 years.

Care and Storage

Caution Clean and maintain the Welch Allyn AED 10 according to instructions.
See

Electrodes

WARNING Follow manufacturer's instructions for use of defibrillation

electrodes.

Caution Properly store and use defibrillation pads.
If a need for repetitive testing arises, wait at least 1 minute for every third
discharge to avoid damaging equipment.
Follow instructions in the Users Manual.
Use only Welch Allyn Part No. 001852 non-rechargeable lithium battery. Use
of any other battery can damage the Welch Allyn AED 10.
Check capacity and change if battery is low. See
page 56.
Battery replacement at 5 years is recommended due to degradation of the
battery chemistry.
"Maintenance"
on page 53.
Do not clean the Welch Allyn AED 10 with alcohol, ketone, or any flammable
agent. Do not autoclave the Welch Allyn AED 10 or attempt to sterilize the
Welch Allyn AED 10 or any of its accessories.
Improper use of defibrillation electrodes may cause the Welch Allyn AED 10
to function improperly or may cause skin burns. Do not use expired, dry
electrodes. Do not reuse disposable electrodes. Only use Welch Allyn
approved electrodes or electrode adapters.
Store electrodes in a cool, dry location (between 60° and 95°F or 15° and
35°C). Do not sterilize the pads, immerse, or clean the electrodes with
alcohol or solvents.
Welch Allyn AED 10 Automatic External Defibrillator
"Maintenance Schedule"
on

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents