User Manual
Program
Restore Defaults
Unit ID
Dept
Change Code
Bit Rate
Utility

Setting the CPR Timer

The CPR timer begins if a "No Shock Advised" decision is made to time the appropriate
length that CPR should be performed.
CPR Timer
120
Save
Cancel
To set the CPR Timer
1.
Access the Supervisor Menu screen. Highlight CPR and press Enter. Select CPR
Timer to display the CPR Timer Menu.
2. Press the Next soft key to scroll through CPR timer lengths.
3. Highlight Save by pressing Enter after the desired time has been selected. Press the
Enter soft key again to confirm the change and return to the CPR Menu.
4. To leave the time as it was originally, press Next when the Save box is highlighted to
select the Cancel box. Press the Enter soft key to return to the CPR Menu.
Available CPR timer values are 15, 30, 60, 90, 120 (default), 150, and 180 seconds.
Note

Setting the CPR Tempo

The CPR Tempo is the rate at which a tone plays when the CPR Timer is running.
CPR Tempo
100
Save
Cancel
To set the CPR Tempo
1.
Access the Supervisor Menu screen. Highlight CPR and press Enter. Select CPR
Tempo to display the CPR Tempo Menu.
2. Press the Next soft key to scroll through CPR tempo values.
Allows the user to update to the current version of Welch Allyn
AED 10 software via the infrared port.
Reset the Welch Allyn AED 10 options to factory defaults.
Set the Welch Allyn AED 10's Unit ID.
Set the Welch Allyn AED 10's Department Name.
Set supervisor password numbers.
Allows the user to change the bit rate of the infrared port.
Opens the Utility Menu
CPR Timer
CPR Timer
150
150
Save
Save
Cancel
Cancel
Cancel
CPR Tempo
CPR Tempo
105
105
Save
Save
Cancel
Cancel
Programming