Specifications
Guidance and manufacturer's declaration—electromagnetic immunity
(IEC 60601-1-2 Table 202)
The Welch Allyn AED 10 is intended for use in the electromagnetic environment specified
below. The customer or the user of the Welch Allyn AED 10 should assure that it is used
in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations on
power supply input lines.
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the ac mains voltage prior to application of the test level.
T
IEC 60601 test level
Compliance level Electromagnetic environment -
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
± 2 kV for power supply
Not applicable
lines
± 1 kV for input/ output
Not applicable
lines
± 1 kV differential mode
Not applicable
± 2 kV common mode
Not applicable
<5% U
(>95% dip in U
)
Not applicable
T
T
for 0.5 cycle
40% U
(60% dip in U
) for
Not applicable
T
T
5 cycles
70% U
(30% dip in U
) for
Not applicable
T
T
25 cycles
<5% U
(>95%dip in U
)
Not applicable
T
T
for 5 sec
3 A/m
10 A/m
Welch Allyn AED 10 Automatic External Defibrillator
guidance
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment