Electromagnetic Compatibility

Category
Radiated Emissions
ESD
Radiated Susceptibility
Magnetic Field Emissions
Magnetic Field Susceptibility
Guidance and manufacturer's declaration—electromagnetic emissions
(IEC 60601-1-2 Table 201)
The Welch Allyn AED 10 is intended for use in the electromagnetic environment specified
below. The customer or the user of the Welch Allyn AED 10 should assure that it is used
in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CSPR 11
Harmonic Emission IEC 6100-3-2
Voltage fluctuations/
flicker emissions IEC 61000-3-3
Medical Electrical Equipment needs special precautions regarding EMC and needs to be
installed and put into service according to EMC information provided in this document.
Standard
Level
EN55011
CISPR11 B
EN61000-4-2
8kV air 6kV contact
EN61000-4-3
10 V/m (20 V/m EN 60601-2-4)
MIL STD RE101
(AAMI DF2 4.3.18.1)
MIL STD RS101
1 Gauss, 47 Hz-1.8 kHz
Compliance
Electromagnetic environment - guidance
Group 1
The Welch Allyn AED 10 uses RF energy only for
its internal function. Therefore, its RF emissions
Class B
are very low and are not likely to cause any
interference in nearby electronic equipment.
Not applicable
Not applicable
Specifications
69