Technical Description; Technical Data - Sirona Heliodent Plus Operating Instructions Manual

X-ray system
Hide thumbs Also See for Heliodent Plus:
Table of Contents

Advertisement

3 Technical description

3.1 Technical data

10
3
Technical description
Technical description
3.1
Technical data
Technical data AL1.5 WW
Technical data
Nominal voltage:
Permissible deviation:
Rated current:
Nominal frequency:
Internal line impedance:
Main building fuse:
Power input during radiation: 1.2 kW
Power input in standby
mode:
Tube voltage:
Tube current:
High-voltage waveform:
High voltage generation
frequency:
Radiation time:
Pulse/pause ratio:
Total filtration of X-ray tube
assembly:
Radiation field:
Dose rate:
Measuring instruments:
Measuring conditions:
Focal spot size as specified
in IEC 336:
Focal spot marking O:
Source-skin distance:
Sirona Dental Systems GmbH
Operating Instructions HELIODENT
120V, 200V– 240V
± 10%
At 120 V: 10 A
At 200 – 240 V: 6 – 5 A
50 Hz / 60 Hz
At 120 V 0.3 ohms
At 200 – 240 V 0.8 Ohm
16 A slow-blow
< 20 W
60 kV / 70 kV switchable
(max. tolerance ± 5 kV)
7 mA (max. tolerance ± 1.4 mA)
DC high frequency
residual ripple value ≤ 4 kV
50 kHz - 70 kHz
0.01 – 3.2 s
(max. tolerance ± 10% +1 ms)
automatic monitoring from 1:1 to 1:60
> 1.5 AI / 70 IEC 60 522
Ø < 60 mm
8.5 mGy/s ±40% at 60 kV
11 mGy/s ±40% at 70 kV
PWT Nomex with an ionization space of
3
1cm
or Unfors mult-o-meter
200 mm focus-meter space
230 V nominal voltage
0.4 mm
FHA 200 mm (8") - standard or 300 mm
(12")
D3507.201.01.06.02 05.2011
PLUS
O
62 14 717 D3507