6 Maintenance

6.2 Inspection and maintenance

30
6.2
Inspection and maintenance
Inspection and maintenance
Inspection and maintenance
Inspection and preventive maintenance must be performed at scheduled
intervals to protect the health and safety of patients, users and other
persons.
Annual inspection
In order to ensure the operational safety and functional reliability of your
product, you as the system owner should check the equipment at regular
intervals (at least once a year) or commission your dental depot to do so.
The information provided in the supplied document "Inspection and
maintenance" should be helpful here.
Maintenance by the service engineer
In addition to the annual check to be carried out by the system owner or
authorized persons, preventive maintenance must be performed after 4,
7 and 10 years, and then at two-year intervals.
The information provided in the supplied document "Inspection and
maintenance" should be helpful here.
Digital image quality
Image quality check
The image quality should be assessed by the system owner at regular
intervals, at least once a year.
On digital image receptor systems, the degree of postprocessing
(brightness or contrast adjustment) that is required in the image
processing software (e.g. SIDEXIS) to produce satisfactory results is
used as an assessment criterion.
Film image quality
For conventional X-rays with film processing, the increase of the
exposure time is used as an assessment criterion.
Assessment criteria
If, after taking into account the patient's anatomy and excluding possible
sources of error such as incorrect patient positioning, this criterion seems
to apply, immediately contact a service engineer to have potential system
faults repaired.
Country-specific requirements
Country-specific requirements
Observe any possible additional country-specific requirements.
Sirona Dental Systems GmbH
Operating Instructions HELIODENT
62 14 717 D3507
D3507.201.01.06.02 05.2011
PLUS