7 Error messages
7.1 List of error messages
32
Error code
Reason and measures
E1 04 51
Safety circuit - the door switch may not be closed
properly
Switch the device OFF and then ON again, check the
door switch
If the error persists, call a service engineer and report the
error code
E3 04 31
Button error - a button may have been pressed during
switch on
Switch the device OFF and then ON again
If the error persists, call a service engineer and report the
error code
Sirona Dental Systems GmbH