SE-1201 Electrocardiograph User Manual
2. When the main screen is displayed, press the MODE key to select the R-R analysis mode.
Press F4 to switch the gain. Press F5 to set the EMG filter or the Lowpass filter.
3. Press the PRINT/STOP key to begin sampling, the sampling time will be displayed on the
main screen. When the sampling time reaches 180s, it begins to analyze and print an ECG
report.
4. It will stop automatically after a complete R-R analysis report is printed, or press the
PRINT/STOP key to stop printing the ECG report.
NOTE: In the R-R analysis mode, you can not set the speed. The constant speed is
25mm/s and the printing speed is 5mm/s, because in the R-R analysis mode, the
ECG wave length is compressed to one fifth of the original wave length.

6.1.5 Review Printing

In the auto or off mode, after you press the Review key,10s ECG data sampled before you press
the key will be printed out.
NOTE: The system will not respond to the REVIEW key unless 10s data has been
sampled on the main screen.

6.2 Copy Printing

In the auto and rhythm modes, pressing the 1mV/COPY key can print the ECG report which was
printed out last time. Pressing the PRINT/STOP key can stop printing the ECG report.
- 51 -