c) The last section of the Local IP item can be set at random.
For the same network transmission,
a) Set the first three sections of the Local IP item to the first three sections of the IP of
the PC.
b) The last section of the Local IP item can be set at random, but it can't be the same as
the last section of the IP of the PC.
4) Set the Gateway item.
Set the first three sections of the Gateway item to the first three sections of the IP of the
electrocardiograph. The last section of the Gateway item must be set to 1.
5) Set the Subnet Mask item to 255.255.255.0.
6) Press Enter to confirm, and then press Esc to return to the main screen.
4. Set Auto Transmission to On in the Transmission Setup window.
5. In the auto or rhythm mode, ECG data will be transmitted through the net cable automatically
after an ECG report is printed out. In the off mode, the sampled ECG data can be saved and
will be transmitted through the net cable automatically after the PRINT/STOP key is pressed,
but it can not be printed.