SE-1201 Electrocardiograph User Manual
Q6: The hint "Memory full! Replace the earliest file?" pops up every time when I operate the
electrocardiograph. What am I supposed to do?
A6: The hint is used to remind you that the amount of stored file reaches 200. You can just make
a choice every time when the hint pops up. Or, you can prevent the display of the pop-up hint
as follows:
Select On from the Replace When Memory Full list box in the File Setup window, if the
amount of stored files reaches 200, the files will replace the earliest ones automatically.
Or, you can just delete several stored files from the electrocardiograph to ensure that the
amount of stored file not to reach 200.
2. Printing Problems
Q1: I was encountered with paper-jam, what was I supposed to do?
A1: If it happened for the first time, it might be the result of an inappropriate placement of the
paper. In this case, please open the recorder casing, pull the paper out of the paper tray, tear
the pages with rumples, and then put the paper in the paper tray again, adjust the position of
the paper carefully and close the casing.
If it happened several times, it might have something to do with your configuration. Please
check the Paper Marker and Paper Style setting, and make sure that the settings match the
loaded paper.
Q2: The hint Paper Error is displayed on the screen, what should I do?
A2: It might be the result of unsuccessful detection of the black markers, first open the recorder
casing so as to clear the error information, and then check whether the black marker is on the
bottom of the paper. Reload the paper in the paper tray. If it doesn't work, change the paper.
If the problem still exists, please contact the manufacturer or the local distributor for further
disposal.
Q3: The hint No Paper is displayed on the screen, what should I do?
A3: Check whether the paper runs out, or the black marker is just facing the black marker
detection window on the thermal printing head, as the following figure shows.
- 109 -
Troubleshooting