13.3.1 Recharge and Replacement of Battery

1) Capacity Identification
The battery capacity can be identified according to the battery symbol in the top right corner
of the LCD screen.
: Full capacity;
: 3/4 capacity;
: 1/4 capacity;
: Low capacity
2) Recharge
SE-1201 electrocardiograph is equipped with the recharge control circuit together with the
battery. When the unit is connected to the mains supply, the battery will be recharged
automatically. Then the battery recharging indicator (
(
) will be lit at the same time. During the recharging course, the symbol
the top right corner of the LCD screen. After the battery is fully recharged, the symbol stops
flashing, and the battery recharging indicator (
Because of the capacity consumption during the storage and transport course, the battery
capacity is not full when it is used for the first time. Battery recharge should be considered
before the first use.
NOTE: The battery will automatically stop charging if you print an ECG report.
Repeated undercharging of the battery will damage the battery and reduce battery life.
3) Replacement
When the useful life of the battery is over, or foul smell and leakage are found, please
contact the manufacturer or the local distributor for replacement.
1. Only qualified service engineers authorized by the manufacturer can open the battery
compartment and replace the battery, and the battery of the same model and
specification provided by the manufacturer must be used.
2. Danger of explosion -- Do not reverse the anode and the cathode when installing the
battery.
) is black.
CAUTION
WARNING