t1: ERR Output respond time ≤5ms
External Trigger Timing Chart
Comparator
________
ERR
Auto Trigger Timing Chart
Comparator
________
ERR
result
(sync)
t1: Trigger pulse width, 0.5ms or more
t2: Delay Time, per setting
t3: Measurement time, the value is Different according to the current
sampling mode, detailed information refer to specification.
t4: Calculation time, 0.18ms, deviation is ±0.02ms
t5: HOLD setting: Holds until the next trigger is detected, PULSE setting:
Remains only for the specified pulse width
result
(sync)
t1: Delay Time, per setting
t2: Measurement time, the value is Different according to the current
sampling mode, detailed information refer to specification.
t3: Calculation time, 0.18ms, deviation is ±0.02ms
t4: EOMoutput pulse width, HOLD setting: Holds until the next trigger is
detected, PULSE setting: Remains only for the specified pulse width
Copyright© Itech
Electronics Co., Ltd.
Function and Features