Use only factory specified spare parts. The manufacturer will not be held
responsible for any accident, incident, or malfunction following a repair
done other than by its Service Center or by an approved repair center.
5.1

Battery Replacement

Before any measurements, make sure that the
when a measurement function is selected. If it is displayed, change all the
batteries. Make sure that none of the terminals are connected and that
the switch is set to OFF before opening the battery compartment.
The battery cover is located on the back of the unit. It can be opened and
closed using a coin or a large screwdriver (1/4-turn captive screw). To
avoid errors, the
the direction in which the 6 x1.5V AA batteries should be inserted. Make
sure that you put the cover back properly.
5.2

Fuse Replacement (1039)

Change the fuse if "FUS" appears on the digital display at power up or
when measuring continuity. Take all necessary precautions when opening
the instrument. It can be opened and closed using a coin or a large screw-
driver (1/4-turn captive screw). Make sure that none of the terminals are
connected and that the switch is set to OFF before opening the back
cover. To avoid any errors, the text "F-0.63A" is written next to the fuse
holder. Make sure that you replace the faulty fuse with fuse of the same
rating and type. Replace and close the cover.
Type of fuse: Fast Acting 0.63A - 600V - 1/4 x 1 1/4" (6.3 x 32mm)
5.3

Cleaning

Disconnect the instrument from any power source. Use a soft cloth
slightly moistened with soapy water. Rinse with a damp cloth and quickly
dry with a dry cloth or forced dry air. Do not use alcohol, solvents or any
hydrocarbon material.
5.4

Storage

If the instrument is to be unused for an extended period (more than 2
Shop for AEMC products online at:
months), remove the batteries and store them separately.