Analog Bargraph
Resolution
Accuracy
Model
Display
0.01 to 39.99MΩ
Resolution
Accuracy
Model
Test Voltage
Open Circuit Voltage
Test Current
Short-Circuit Current
The residual voltage present on the terminals when the yellow button is
released is automatically discharged through the instrument at a rate of
20s/µF (1039) and 1.5s/µF (1035).
3.1.3
Resistance
Measurement Range: 0 to 40kΩ (1039)
Model
Ranges Automatic
Resolution
Accuracy
Test Current
Open Circuit Voltage
3.1.4
Capacitance (1039)
Measurement Range: 0 to 4000nF
Range
Resolution
Accuracy
Measurement Current
Shop for AEMC products online at:
Digital Analog Megohmmeter Models 1035/1039
80kΩ to 1.1GΩ
10 segments per 10-unit interval
5% R ± 1 segment
1039 and 1035
40.0 to 399.9MΩ
10kΩ
3% R ± 5cts
1039 and 1035
50V
<75V
≥1mA for R
≤ 50kΩ
0 to 400kΩ (1035)
0.0 to 399.9Ω 400 to 3999Ω 4.00 to 39.99kΩ 40.0 to 399.9kΩ
0.1Ω
±3% R ± 2cts
600µA
www.
400 to 3999MΩ
100kΩ
3% R ± 2cts
100V
<150V
≥1mA for R ≤
100kΩ
≤ 3mA
1039 and 1035
1Ω
7V ≤ V open ≤ 9V
0.00 to 399.9nF
0.1nF
±3% ± 2cts
1mA
ShopAEMC
1035
1MΩ
3% R ± 2cts
1035
250V
<375V
≥1mA for R
≤ 250kΩ
10Ω
±3% R ± 1ct
6µA
400 to 3999nF
1nF
±3% ± 1ct
100µA
.ca
1.877.766.5412
4.00 to 20GΩ
10MΩ
500V
<750V
≥1mA for R ≤
500kΩ
1035
100Ω