Model 1039: 10kΩ at 50V; 20kΩ at 100V
Model 1035: 10kΩ at 50V; 20kΩ at 100V; 50kΩ at 250V; 100kΩ at
500V
When the measurement is less than 70kΩ, only the ◄ symbol lights up
on the left-hand side of the bargraph.
To measure high insulation values, you are advised to use the "G" guard
terminal to remove the influence of superficial leak currents and "hand
capacitance".
2.6.3

Resistance

Resistance measurement corresponds to the 40kΩ position of the switch
on the 1039 and the 400kΩ position on the 1035. The measurement is
indicated on the digital measurement display, accompanied by the Ω or
kΩ symbol if necessary. The instrument indicates if the value measured is
outside its measurement range. If the resistance is greater than 40kΩ or
400kΩ, depending on the instrument, the OL symbol is displayed on the
digital measurement display. A programmed threshold may trigger an
alarm (see pg 18).
2.6.4

Capacitance (1039)

Capacitance measurement corresponds to the 4000nF position of the
switch. The measurement is shown on the display accompanied by the
nF symbol. The length of the telephone line measured is indicated in km
on the digital threshold display, according to the programmed capacitance
per unit length.
If the leads are set up in a short circuit, the instrument indica-
tions OL on the digital measurement display and the distance is
- - - km.
If the capacitance measurement is greater than 4000nF, the
OL symbol is displayed on the digital measurement display.
If the distance measurement is greater than 80km, the OL
symbol is displayed on the digital measurement display.
If the leads are not connected, the digital measurement display
indicates 0.00.