Chapter 3: Testing Performance

Performing Self-Test and Calibration

1 Warm up the oscilloscope for at least 30 minutes.
2 Perform self tests.
a Pull down the Utilities menu and select Self Test.
b Select Scope Self Test from the Self Test list.
c Click Start Test to start the self test procedure.
If any of the self tests fail, ensure that the failure is diagnosed and repaired before calibrating
and testing performance.
3 Perform calibration. See "To run the self calibration" on page 14.
Performing Self-Test and
Calibration
19