Keysight 9000 Series Service Manual page 89

Table of Contents

Advertisement

as shown in Figure 17.
Figure 17
7 There are two tabs on the edge closest to the touch screen that need to be disengaged. To disengage the
one near the top of the oscilloscope, look down the edge from the top and press in on the tab (meaning
press straight towards the bottom of the oscilloscope) with a flathead screwdriver while pulling up on
the plastic keyboard cover / keyboard. The plastic piece and keyboard will not come all the way off
because you need to repeat this procedure for the bottom tab (look along the inner edge from the bottom,
push in on the bottom tab (meaning push towards the top of the oscilloscope), and then pull up on the
plastic keyboard cover / front panel keyboard). Figure 18 shows a picture of performing this procedure
on the upper tab.
Figure 18
Plastic piece
covering
front panel
keyboard
Top inner tab that
you need to disengage
Push flathead screwdriver
straight back (toward
the bottom of the
oscilloscope) and pull up like this picture shows
to disengage this tab.
Chapter 5: Replacing Assemblies
front panel bezel, front panel keyboard, touch screen, and Autoprobe board
Disengage three tabs
To remove and replace the
Top of the oscilloscope
89

Advertisement

Table of Contents
loading

Table of Contents