To remove and replace the front panel bezel, front panel keyboard, touch
screen, and Autoprobe board
Disconnect the power cable and remove the handle and rear cover as described above in the
1
section on removing the handle and rear cover.
2 Position the oscilloscope upright and disengage the series of tabs around the outside of the front bezel
as shown in Figure 13.
3 With the tabs disengaged, pull the front bezel off as shown in Figure 14.
Figure 13
Figure 14
Chapter 5: Replacing Assemblies
front panel bezel, front panel keyboard, touch screen, and Autoprobe board
To remove and replace the