Keysight 9000 Series Service Manual page 51

Table of Contents

Advertisement

Test
Limits
Time Scale
Limit
Accuracy
Limit is based on years since oscilloscope's last
calibration
TSA = ±(0.4 + 0.5(Years Since Calibration)) ppm
Chapter 3: Testing Performance
Results
Calculated
Measured time
TSA Spec
scale error
___________
___________
Performance Test Record
Pass/Fail
___________
51

Advertisement

Table of Contents
loading

Table of Contents