A Quick Help system is built into the oscilloscope. Press and hold any key to display Quick Help. Complete instructions for using the quick help system are given in “Access the Built-In Quick Help" on page 28. For more information about InfiniiVision oscilloscopes, see: www.keysight.com/find/scope Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
When using licensed serial bus • “Serial Bus Decode/Trigger" on page 73 decode and triggering features, see: When saving, recalling, or printing, • “Save/Recall (Setups, Screens, Data)" on page 79 see: • “Print (Screens)" on page 82 Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 6
[Key1], then pressing Softkey2, then pressing Softkey3 are abbreviated as follows: Press [Key1]> Softkey2 > Softkey3. The keys may be a front panel [Key] or a Softkey. Softkeys are the six keys located directly below the oscilloscope display. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Front Panel Overlays for Different Languages / 24 Learn the Rear Panel Connectors / 25 Learn the Oscilloscope Display / 26 Access the Built-In Quick Help / 28 Quick Reference Running, Stopping, and Making Single Acquisitions (Run Control) / 30 Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 8
Trigger Mode, Coupling, Reject, Holdoff / 49 External Trigger Input / 51 Maximum voltage at oscilloscope external trigger input / 51 Acquisition Control / 52 Selecting the Acquisition Mode / 52 Overview of Sampling / 53 Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 9
Configuring the [Quick Action] Key / 85 Specifications and Characteristics / 86 Environmental Conditions / 87 Declaration of Conformity / 87 Probes and Accessories / 88 Software and Firmware Updates / 89 Acknowledgements / 90 Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 10
Index Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
• Verify that you received the following items and any optional accessories you may have ordered: • InfiniiVision 1000 X-Series oscilloscope. • Power cord (country of origin determines specific type). • Two oscilloscope probes. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
3 Press the power switch. The power switch is located on the lower left corner of the front panel. The oscilloscope will perform a self-test and will be operational in a few seconds. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Do not negate the protective action of the ground connection to the oscilloscope. The WARNING oscilloscope must remain grounded through its power cord. Defeating the ground creates an electric shock hazard. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
1 Connect an oscilloscope probe from channel 1 to the Demo, Probe Comp terminal on the front panel. 2 Connect the probe's ground lead to the ground terminal (next to the Demo terminal). Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
In the Save/Recall menu, there are also options for restoring the complete factory settings or performing a secure erase (see “Save/Recall (Setups, Screens, Data)" on page 79). Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
“Compensate Passive Probes" on page 18. If you do not see the waveform, make sure the probe is connected securely to the front panel channel input BNC and to the Demo/Probe Comp terminal. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
BNC connector. On other probes (like the N2862/63/90 probes), the trimmer capacitor is a yellow adjustment on the probe tip. Perfectly compensated Over compensated Under compensated 7 Connect probes to all other oscilloscope channels. 8 Repeat the procedure for each channel. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
[Intensity] key Press the key to illuminate it. When illuminated, turn the Entry knob to adjust waveform intensity. You can vary the intensity control to bring out signal detail, much like an analog oscilloscope. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 20
52). • Zoom key — Press the zoom key to split the oscilloscope display into Normal and Zoom sections without opening the Acquire menu. For more information see “Horizontal Controls" on page 31. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
See “Configuring the [Quick Action] Key" page 85. • [Save to USB] key — Press this key to perform a quick save to a USB storage device. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 22
• Enable serial bus decodes. See also “Serial Bus Decode/Trigger" on page 73. Ext Trig input External trigger input BNC connector. See “External Trigger Input" on page 51 for an explanation of this feature. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 23
18. With certain licensed features, the oscilloscope can also output demo or training signals on this terminal. • Ground terminal — Use the ground terminal for oscilloscope probes connected to the Demo/Probe Comp terminal. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
To install a front panel overlay: 1 Gently pull on the front panel knobs to remove them. 2 Insert the overlay's side tabs into the slots on the front panel. 3 Reinstall the front panel knobs. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
This is where you can attach a Kensington lock for securing the instrument. hole USB Device This port is for connecting the oscilloscope to a host PC. You can issue remote commands from a port host PC to the oscilloscope via the USB device port. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Each analog channel's information appears in a different color. Signal detail is displayed using 256 levels of intensity. For more information about display modes see “Display Settings" on page 45. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 27
When measurements or cursors are turned on, this area contains automatic measurement and cursor results. When measurements are turned off, this area displays additional status information describing channel offset and other configuration parameters. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
To select the user interface and Quick Help language: interface and 1 Press [Help], then press the Language softkey. Quick Help 2 Turn the Entry knob until the desired language is selected. language Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
[Run/Stop] (the key is red when stopped) Single acquisition [Single] (the key is yellow until the oscilloscope triggers) If the oscilloscope does not trigger, you can press [Force Trigger] to trigger on anything and make a single acquisition. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
The following figure shows the Acquire menu which appears after pressing the [Acquire] key. Trigger Time Delay Time/ Trigger Trigger level point reference time source or threshold Normal time mode XY or Roll mode Zoomed time base Time reference Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Zoom The Zoom window is a magnified portion of the normal time/div window. To turn on (or off) Zoom, press the zoom key (or press the [Acquire] key and then the Zoom softkey). Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 33
Zoom window window window when the Horizontal position knob is turned Normal window Signal anomaly expanded in zoom window Zoom window Select Zoom Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Quick Reference Vertical Controls Vertical Knobs and Keys Vertical Softkey Controls The following figure shows the Channel 1 menu that appears after pressing the [1] channel key. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 35
Note that Channel Coupling is independent of Trigger Coupling. To change trigger coupling see “Trigger Mode, Coupling, Reject, Holdoff" on page 49. Channel bandwidth [1/2] > BW Limit limit Vertical scale fine [1/2] > Fine adjustment Channel Invert [1/2] > Invert Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Entry knob Probe check [1/2] > Probe > Probe Check Guides you through the process of compensating passive probes (such as the N2140A, N2142A, N2862A/B, N2863A/B, N2889A, N2890A, 10073C, 10074C, or 1165A probes). Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Place the desired part of the waveform at the center of the screen and decrease frequency span to increase the display resolution. As frequency span is decreased, the number of points shown is reduced, and the display is magnified. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 39
• Adjust frequency span for better cursor placement. • Return to the Cursors Menu to fine tune the X cursor. For more information on the use of FFTs please refer to Keysight Application Note 243, The Fundamentals of Signal Analysis at http://literature.cdn.keysight.com/litweb/pdf/5952-8898E.pdf.
FFT resolution of 1.91 Hz. The displayed FFT spectrum waveform shows the components of the input signal above the Nyquist frequency to be mirrored (aliased) on the display and reflected off the right edge. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
FFT. The FFT Menu provides four windows: Hanning, Flat Top, Rectangular, and Blackman-Harris. For more information on leakage, see Keysight Application Note 243, The Fundamentals of Signal Analysis at http://literature.cdn.keysight.com/litweb/pdf/5952-8898E.pdf.
[Math] > Auto Setup Span/center [Math] > More > Span [Math] > More > Center Window function [Math] > More > Window (Hanning, Flat Top, Rectangular, Blackman Harris, see also “FFT Spectral Leakage" on page 41) Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
A scale unit of U (undefined) will be displayed for math functions when two source channels are used and they are set to dissimilar units and the combination of units cannot be resolved. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
See also “To load a list of labels from a text file you create" on page 46. Label library reset [Utility] > Options > Preferences > Defaul t Library Annotations [Display] > Annotation > Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
2 Name the file labellist.txt and save it on a USB mass storage device such as a thumb drive. 3 Load the list into the oscilloscope using the File Explorer (press [Utility] > File Explorer). Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 47
When you reset the label library list (see next topic), all of your custom labels will be deleted, and the label list will be returned to its factory configuration. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
In addition to the edge trigger type, you can set up triggers on pulse widths and video signals. In the DSOX1000-Series oscilloscopes, you can also set up triggers on patterns, rising and falling edge transition times, and setup and hold violations. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
NOTE: Many video signals are produced from 75 sources. To provide correct matching to these Ω sources, a 75 terminator (such as a Keysight 11094B) should be connected to the oscilloscope input. Pattern trigger [Trigger] > Trigger Type, Pattern Rise/fall edge [Trigger] >...
Page 50
[Trigger] > Coupling (DC, AC, LF Reject, TV/Video) NOTE: Trigger coupling is independent of channel coupling (see “Vertical Controls" on page 34). Trigger noise reject [Trigger] > Noise Rej Trigger high frequency [Trigger] > HF Reject reject Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
[External] > Range, Entry knob For DSOX1000-Series oscilloscopes only. On EDUX1000-Series oscilloscopes, the range is fixed at 8 V when you are using a 1:1 probe. External trigger [External] > Position, Entry knob waveform position Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
[Acquire] > Acq Mode, High Resolution acquisition mode At slower time/div settings, all samples in the effective sample period are averaged and the average value is stored. Use this mode for reducing random noise. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
) = folding frequency Aliasing Aliasing occurs when signals are under-sampled (f < 2f ). Aliasing is the signal distortion caused by low frequencies falsely reconstructed from an insufficient number of sample points. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 54
At the oscilloscope bandwidth, sampling theory says the required sample rate is f = 2f . However, the theory assumes there are no frequency components above in this case) and it requires a system with an ideal brick-wall frequency response. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 55
(square waves are made up of sine waves at the fundamental frequency and an infinite number of odd harmonics), and typically, for 500 MHz bandwidths and below, oscilloscopes have a Gaussian frequency response. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 56
See Also Evaluating Oscilloscope Sample Rates vs. Sampling Fidelity: How to Make the Most Accurate Digital Measurements, Keysight Application Note 1587 (http://literature.cdn.keysight.com/litweb/pdf/5989-5732EN.pdf) Oscilloscope Rise Time Closely related to an oscilloscope's bandwidth specification is its rise time specification.
= 1.9 x f knee See Also Choosing an Oscilloscope with the Right Bandwidth for your Application, Keysight Application Note 1588 (http://literature.cdn.keysight.com/litweb/pdf/5989-5733EN.pdf) Memory Depth and Sample Rate The number of points of oscilloscope memory is fixed, and there is a maximum sample rate associated with oscilloscope's analog-to-digital converter;...
Page 58
Likewise, when storing 50 ms of data in 50,000 points of memory, the actual sample rate is 1 MSa/s. The actual sample rate is displayed in the right-side information area. The oscilloscope achieves the actual sample rate by throwing away (decimating) unneeded samples. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
FFT where the values are relative to 0 dB. In XY horizontal mode, the Y cursors display channel 2 values (Volts or Amps). Cursor Knobs and Keys Cursor Softkey Controls Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 60
[Cursors] > Mode, Binary Hex cursors mode [Cursors] > Mode, Hex Cursors X units [Cursors] > Units > X Units (Seconds, Hz, Phase, Ratio) Cursors Y units [Cursors] > Units > Y Units (Base, Ratio) Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
All measurements are available for analog channel waveforms. All measurements except Counter are available for math waveforms other than FFT. A limited set of measurements is available for math FFT waveforms. Use the cursors to make other measurements on FFT. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Creating/Editing Mask Files A mask file contains the following sections: • Mask File Identifier. • Mask Title. • Mask Violation Regions. • Oscilloscope Setup Information. Mask File Identifier The Mask File Identifier is MASK_FILE_548XX. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 64
Similarly, the vertices in Region 2 describe the line that forms the top of the bottom part of the mask. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 65
The mask scaling controls how the normalized vectors are interpreted. This in turn controls how the mask is drawn on the display. The remote programming commands that control mask scaling are: :MTES:SCAL:BIND 0 :MTES:SCAL:X1 -400.000E-06 :MTES:SCAL:XDEL +800.000E-06 Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 68
4 regions. When there are 4 regions in a vertical column, one region must be tied to the top (using the MAX y value) and one must be tied to the bottom (using the MIN y value). Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 69
"# of Tests". The mask file allows greater resolution than the 200 X 640 database. Some quantization of data occurs to reduce the mask file data for display on-screen. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
The Frequency mode requires the Edge or Pulse Width trigger type, and the DVM source and the trigger source must be the same analog channel. Auto range [Analyze] > Auto Range enable/disable Auto range can be used when the DVM input channel is not used in oscilloscope triggering. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
[Analyze] > Chart > Autoscale phase plots View measured gain [Analyze] > Move Marker, Entry knob and phase values Save data to USB [Save/Recall] > Save > Format, Frequency Response Analysis data (*.csv) storage device Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
To order serial decode licenses, go to www.keysight.com and search for the product number (for example, DSOX1AUTO) or contact your local Keysight Technologies representative (see www.keysight.com/find/contactus). Triggering on When triggering on a slow serial signal (for example, I2C, SPI, CAN, LIN, etc.) it...
Entry knob (Rx, Tx, CAN_H, CAN_L, Differential (L-H), Differential type/polarity (H-L)) CAN counters [Bus] > Reset CAN Counters Counters run even when the oscilloscope is stopped (not acquiring data). When an overflow condition occurs, the counter displays OVERFLOW. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
For triggers where you can specify address or data values, there are additional softkeys you can use to enter the values. LIN Decode/Trigger On DSOX1000-Series oscilloscopes, the LIN serial decode and triggering option can be enabled with the AUTO license. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
SPI Decode/Trigger On DSOX1000-Series oscilloscopes, the SPI serial decode and triggering option can be enabled with the EMBD license. When interpreting the decode waveforms, see the Mode softkey built-in help. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
UART Bus Config menu. In this menu, you can select the oscilloscope source channels probing the Rx and Tx signals and the appropriate threshold voltage for each. UART/RS232 number [Bus] > Bus Config > # Bits, Entry knob (5, 6, 7, 8, 9) of bits Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 78
For triggers where you can specify data values, there are additional softkeys for specifying the data comparison operator, the data value, the data value base (Hex or ASCII), and the Nth frame burst count. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Single acquisitions use full memory. • Whether only one channel of a pair is turned on. (Channels 1 and 2 are one pair.) Acquisition memory is divided among the channels in a pair. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 81
For example: if the Length is set to 1000, and you are displaying a record that is 5000 data points in length, four of each five data points will be decimated, creating an output file 1000 data points in length. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
To print the current screen once the printer is configured (and Quick Print is selected as the quick action), simply press the [Quick Action] key. For the most up-to-date listing of printers that are compatible with the InfiniiVision oscilloscopes, please visit www.keysight.com/find/InfiniiVision-printers. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
[Utility] > Service > Start User Calibration Service task support [Utility] > Service > See also the Keysight InfiniiVision 1000 X-Series Oscilloscopes Service Guide. The service guide also tells you how to: • Clean the oscilloscope • Check warranty and extended services status •...
Quick Reference Specifications and Characteristics For up-to-date specifications and characteristics on the 1000 X-Series oscilloscopes, see the data sheet at: www.keysight.com/find/1000X-Series Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive pollution occurs which becomes conductive due to condensation which is expected. Example: Sheltered outdoor environment. Declaration of Conformity For Declarations of Conformity for Keysight products, go to: www.keysight.com/go/conformity Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
InfiniiVision Oscilloscope Probes and Accessories Selection Guide Data Sheet (5968-8153EN) • For compatibility information, manuals, application notes, data sheets, selection guides, SPICE models, and more for oscilloscope probes, see the Probe Resource Center at: www.keysight.com/find/PRC Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Quick Reference Software and Firmware Updates From time to time Keysight Technologies releases software and firmware updates for its products. To search for firmware updates for your oscilloscope, direct your web browser to www.keysight.com/find/1000X-Series-sw. To view the currently installed software and firmware press [Help] > About Oscilloscope.
Quick Reference Acknowledgements HDF5 Reference Waveform files use HDF5. HDF5 was developed by The HDF Group and by the National Center for Supercomputing Applications at the University of Illinois at Urbana-Champaign. Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 91
Bode plot, frequency response color printing, analog channel, probe attenuation, analysis, color screen image, analog channel, setup, brick-wall frequency response, compensate passive probes, 18, analog filters, adjusting, brightness of waveforms, Conformity, Declaration of, Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...
Page 92
(FRA), degrees, FFT (Phase) vertical units, frequency response analysis data, Delay measurement, factory default settings, save, delay time, Fall time measurement, frequency, Nyquist, delete file, fast debug AutoScale, Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide...