Figure 27
The touch screen is
mounted on the other
side of this sheet metal
Main oscilloscope
body
Figure 28
Also disconnect this
display cable
These are the three
cables mentioned in
the figure immediately
above that need
disconnected before
you remove the
touch screen display
16 Disconnect the touch screen cable (the one with the blue connector) and the three display cables (the
ones with the white connectors) as shown in Figure 27 and Figure 28.
17 Remove the four T10 screws that attach the touch screen to the sheet metal (there are two on each side
of the touch screen - right and left). See Figure 29. When reattaching these screws, tighten to 5 in-lbs.
Chapter 5: Replacing Assemblies
front panel bezel, front panel keyboard, touch screen, and Autoprobe board
Disconnect this cable (the one with the blue connector) because it
is connected to the touch screen
To remove and replace the
Remove these two
display cables
(white connectors)
from the inverter
board because they
are connected to the
touch screen display
95