Chapter 3: Testing Performance
d Set the horizontal scale to 16 ns/div (to display 8 cycles of a 50 MHz waveform).
Click here and enter 16E-9
e Pull down the Setup menu, select Acquisition, and set up the acquisition parameters as follows:
Sin(x)/x Interpolation filter = Auto
Analog Averaging = Enabled with # of Averages set to 16 samples
Analog Memory Depth = Automatic
Analog Sampling rate = Maximum (check the Manual button, click the field displaying the value,
and then click the Max button in the dialog box that appears)
f Pull down the Measure menu, select Voltage, and then select V rms.
Analog Bandwidth -
Maximum Frequency Check
37