To Set The Calibration Factors After Replacing The Acquisition Board - Keysight 9000 Series Service Manual

Table of Contents

Advertisement

Chapter 5: Replacing Assemblies

To set the calibration factors after replacing the acquisition board

To set the calibration factors after replacing the acquisition board
The following procedure must be performed after replacing the acquisition board. This procedure needs
to be performed only once after the acquisition board is replaced.
Let the oscilloscope warm up before testing
The oscilloscope under test must be warmed up (with the oscilloscope application running) for at
least 30 minutes prior to the start of any performance test.
Equipment Required
Description
Digital Multimeter
10 MHz Signal
Source (required for
time scale
calibration)
Cable Assembly
Cable Assembly
Cable Assembly
Adapter
Adapter
Adapter (Qty. 2)
Adapter
Perform self tests.
1
a Exit the oscilloscope application.
b From a command prompt, run the oscilloscope software using the following syntax:
c:\scope\bin\agscope.exe -service
While the oscilloscope is warming up, run the self-test to ensure that the hardware is functioning
properly. To run the self-test:
c Pull down the Utilities menu and select Self Test.
d Select Scope Self Test from the Self Test list.
e Click on Start Self Test to start the self test procedure.
If any of the self-tests fail, ensure that the failure is diagnosed and repaired before calibrating
and testing performance.
2 Performance calibration.
After the warm up period, calibrate the oscilloscope.
a Connect the RS-232 cable to the Digital Multimeter RS-232 port and to the RS-232 port on the
oscilloscope.
b Set Digital Multimeter as follows:
Interface: RS-232
Baud Rate: 9600 Baud
Parity: None: 8 bits
Language: SCPI
108
Critical Specifications
No substitute
Frequency accuracy better than 0.4 ppm
No substitute
RS-232 (f)(f)
Ω
50
characteristic impedance BNC (m) connectors ~ 36 inches
(91 cm) to 48 inches (122 cm) long
BNC Barrel (f)(f)
BNC shorting cap
No substitute
BNC (f) to dual banana
Recommended
Keysight Model/
Part Numbers
34401A
53131A with Opt. 010
54855-61620
34398A
8120-1840
1250-0080
1250-0929
54855-67604
1251-2277

Advertisement