To run the self calibration

Let the Oscilloscope Warm Up Before Adjusting
Warm up the oscilloscope for 30 minutes before starting calibration procedure. Failure to allow
warm up may result in inaccurate calibration.
The self calibration uses signals generated in the oscilloscope to calibrate channel sensitivity, offsets,
and trigger parameters. You should run the self calibration
• yearly, or according to your periodic needs,
• when you replace the acquisition assembly or acquisition hybrids,
• when you replace the hard drive or any other assembly,
• when the oscilloscope's operating temperature (after the 30-minute warm-up period) is more than
±5 °C different from that of the last calibration.
Equipment Required
Equipment
Ω
50
BNC Cable
10 MHz Signal Source (required for time scale
calibration)
MSO calibration fixture that ships with the
oscilloscope (MSO models only)
MSO cable that ships with the oscilloscope
(MSO models only)
Self calibration
Calibration time
It will take approximately 1 hour to run the self calibration on the oscilloscope, including the time
required to change cables from channel to channel.
1 Let the oscilloscope warm up before running the self calibration.
The self calibration should only be done after the oscilloscope has run for 30 minutes at ambient
temperature with the cover installed. Calibration of an oscilloscope that has not warmed up may result
in an inaccurate calibration.
Pull down the Utilities menu and select Calibration.
2
Click the check box to clear the Cal Memory Protect condition.
3
You cannot run self calibration if this box is checked. See Figure 1.
14
Critical Specifications
Ω
50
characteristic impedance
BNC cable
Frequency accuracy better
than 0.4 ppm
Keysight Part Number
Keysight 53131A with
Opt. 010
Keysight 54904-61615