Run A User Calibration; Verify System Performance - Keysight Infiniium S Series Service Manual

Hide thumbs Also See for Infiniium S Series:
Table of Contents

Advertisement

To verify correct touch screen operation:
1
2
If the touch screen is not working properly, go to "Touch Screen Troubleshooting"
in this chapter.

Run a user calibration

1
2

Verify system performance

After you have verified the basic operation of the oscilloscope, you need to verify
that it meets all warranted specifications by following the procedures in the
"Testing
S-Series Oscilloscopes Service Guide
Click Utilities > Self Test....
Select Touch Screen Test from the Available Self Tests drop-down list box, then
click Start and follow the on-screen instructions.
Complete a user calibration by following the procedures in
If the calibration test fails, replace the acquisition board.
Performance" chapter.
4
Troubleshooting
Chapter
2.
57