Calibration

To run the self calibration

Let the Oscilloscope Warm Up Before Adjusting
CAUTION
Warm up the oscilloscope for 30 minutes before starting calibration procedure.
Failure to allow warm up may result in inaccurate calibration.
The self calibration uses signals generated in the oscilloscope to calibrate channel
sensitivity, offsets, and trigger parameters. You should run the self calibration
• yearly, or according to your periodic needs,
• when you replace the acquisition assembly or acquisition hybrids,
• when you replace the hard drive or any other assembly,
• when the oscilloscope's operating temperature (after the 30-minute warm-up
Table 2 Equipment Required
Equipment
50 Ω BNC Cable
10 MHz Signal Source
(required for time scale
calibration)
MSO calibration fixture that
ships with the oscilloscope
(MSO models only)
MSO cable that ships with the
oscilloscope (MSO models
only)
Self calibration
Calibration time
N OTE
It will take approximately 1 hour to run the self calibration on the oscilloscope, including the
time required to change cables from channel to channel.
1
2
20
period) is more than ±5 °C different from that of the last calibration.
Let the oscilloscope warm up before running the self calibration.
The self calibration should only be done after the oscilloscope has run for 30
minutes at ambient temperature with the cover installed. Calibration of an
oscilloscope that has not warmed up may result in an inaccurate calibration.
Pull down the Utilities menu and select Calibration.
Critical Specifications
50 Ω characteristic impedance
BNC cable
Frequency accuracy better
than 0.4 ppm
Infiniium 9000 Series Oscilloscopes Service Guide
Keysight Part Number
Keysight 53131A with Opt. 010
Keysight 54904-61615