To set the calibration factors after replacing the acquisition board

The following procedure must be performed after replacing the acquisition board.
This procedure needs to be performed only once after the acquisition board is
replaced.
Let the oscilloscope warm up before testing
N OTE
The oscilloscope under test must be warmed up (with the oscilloscope application running) for
at least 30 minutes prior to the start of any performance test.
Equipment
Required
Description
Digital Multimeter
10 MHz Signal Source
(required for time
scale calibration)
Cable Assembly
Cable Assembly
Cable Assembly
Adapter
Adapter
Adapter (Qty. 2)
Adapter
1
Infiniium 9000 Series Oscilloscopes Service Guide
Critical Specifications
No substitute
Frequency accuracy better than 0.4 ppm
No substitute
RS-232 (f)(f)
50 Ω characteristic impedance BNC (m)
connectors ~ 36 inches (91 cm) to 48 inches
(122 cm) long
BNC Barrel (f)(f)
BNC shorting cap
No substitute
BNC (f) to dual banana
Perform self tests.
Exit the oscilloscope application.
a
From a command prompt, run the oscilloscope software using the following
b
syntax: c:\scope\bin\agscope.exe -service
While the oscilloscope is warming up, run the self-test to ensure that the
hardware is functioning properly. To run the self-test:
Pull down the Utilities menu and select Self Test.
c
Select Scope Self Test from the Self Test list.
d
Click on Start Self Test to start the self test procedure.
e
5
Replacing Assemblies
Recommended
Keysight Model/Part
Numbers
34401A
53131A with Opt. 010
54855-61620
34398A
8120-1840
1250-0080
1250-0929
54855-67604
1251-2277
119