3.0 Connecting conditions
3.1 Required connections for all units of ATMOS
Flexible multi-port distributors may not be used as a power supply for the ATMOS
Servant vision or for an optional HF or radiofrequency surgical device.
Unit / Device
ATMOS
S 61 Servant ENT workstation
®
ATMOS
S 61 Servant vision
®
ATMOS
S 61 Servant instruments
®
Water separating system (WTA)
Monitor
Patient chair
An adequate number of socket outlets with earthing contact should be mounted for possible connection of further
electrically operated units which may be installed (e.g. installation of an electrically operated ATMOS
water separating system, camera, monitor, etc.)

3.2 Connection to electrical power line

Prerequisites
• Installation acc. to IEC 60346-7-710: earth leakage circuit breaker (FI-circuit breaker) with rated leakage current <0.03 A
• Connection of the power supply cable of the ATMOS
meters, preferably left (fig.).
• The supply circuit must be separated from other devices e.g. PCs etc.
• If isolating transformers are used then isolation monitoring must be integrated in the isolating transformer.
• Maximum power consumption:
- ATMOS
S 61 Servant workstation
®
- ATMOS
S 61 Servant vision
®
- ATMOS
S 61 Servant instruments
®
Flexible multi-port distributors may not be used as a power supply for the ATMOS
workstation.
i
First start-up
Prior to shipment each ATMOS
to make sure that the appliance is working safely after transport and installation, the following points should be observed: The user
should put the appliance into operation only if
1. a functional test on the appliance at the place of operation has been carried out.
2. the operating instructions have been read and noticed.
Following transportation at low temperatures the appliance must be held for up to fours hours at ambient temperature before first
start-up. When the appliance has not been acclimatised the formation of condensation water is possible and a malfunction might be
the result.
i
Connect the unit to a mains socket with the supplied IEC power connector. Please observe country-specific voltage!
The ATMOS
S 61 Servant instruments has no mains switch for switching on the devices. Each optional available electrical device
®
has its own mains switch.
For separation from the power supply please disconnect the power plug!
Maximum required connections
1 x earthing contact socket outlet
1 x earthing contact socket outlet (basic version) or
1 x fixed connection for the integrated camera or stroboscope LED
3 x earthing contact socket outlet
1 x earthing contact socket outlet
1 x earthing contact socket outlet
1 x earthing contact socket outlet
S 61 Servant instruments to a safety socket outlet near the device, max. 3
®
2.300 VA
200 VA
250 VA
S 61 Servant instruments is being inspected by the manufacturer for function and safety. In order
®
S 61 Servant
®
S 61 Servant workstation, ATMOS
®
S 61 Servant vision or ATMOS
®
S 61
®
patient chair,
®
S 61 Servant
®