10.0 Notes on EMC

i
• Medical electrical equipment is subject to special precautions with regard to EMC and must be installed acc. to following EMC
notes.
• Portable and mobile HF communication facilities can influence medical electrical equipment.
• The use of other accessories, other converters and cables than stated may lead to an increased emission or a reduced
interference immunity of the equipment or system.
10.1 Guidelines and Manufacturer´s Declaration - Emissions
The ATMOS
S 61 Servant instruments is intended for use in the electromagnetic environment specified below. The customer or
®
user of the ATMOS
S 61 Servant instruments should ensure that it is used in such an environment.
®
Emissions Test
RF Emissions acc.to CISPR 11
RF Emissions according to CISPR
11
Harmonic emissions according to
IEC 61000-3-2
Voltage fluctuations/flicker according
to IEC 61000-3-3
i
The device may not be used directly next to other devices or piled up with other devices. If operation next to or piled with other
devices is necessary, please watch the device to check its intended operation in this arrangement.
10.2 Guidelines and Manufacturer's Declaration - Immunity
The ATMOS
S 61 Servant instruments is intended for use in the electromagnetic environment specified below. The customer or
®
user of the ATMOS
S 61 Servant instruments should ensure that it is used in such an environment.
®
Immunity Test
Electrostatic discharge
(ESD) according to IEC
61000-4-2
Fast electrical transient/
burst IEC 61000-4-4
Surges IEC 61000-4-5
Magnetic field at power
frequency 50/60 Hz acc.
to IEC 61000-4-8
Compliance
Group 1
Class B
Inapplicable
Inapplicable
IEC 60601- Test Level
Compliance Level
± 6 kV Contact
± 6 kV Contact
± 8 kV Air
± 8 kV Air
± 2 kV Mains
± 2 kV Mains
± 1 kV I/Os
Inapplicable
± 1 kV differential mode
± 2 kV differential mode
± 1 kV common-mode
± 1 kV common-mode
3 A/m
Inapplicable
Electromagnetic Environment - Guidance
The ATMOS
S 61 Servant instruments uses RF energy only
®
for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic
equipment.
The ATMOS
S 61 Servant instruments is suitable for use in all
®
establishments, including domestic, and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Electromagnetic Environment - Guidance
Floors should be wood, concrete, or ceramic
tile. If floors are synthetic, the relative humidity
should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Power frequency magnetic fields should be that
of a typical commercial or hospital environment.