3.0 Assembly: Connecting conditions

These connection conditions refer to the overall unit with maximum equipment.
3.1 Required connections for all units of ATMOS
Flexible multi-port distributors may not be used as a power supply for the ATMOS
61 Servant vision or for an optional HF or radiofrequency surgical device.
Unit / Device
ATMOS
S 61 Servant ENT workstation
®
ATMOS
S 61 Servant vision
®
ATMOS
S 61 Servant instruments
®
Water separating system (WTA)
Monitor
Patient chair
An adequate number of socket outlets with earthing contact should be mounted for possible connection of further
electrically operated units which may be installed (e.g. installation of an electrically operated ATMOS
water separating system, camera, monitor, etc.)

3.2 Connection to electrical power line

Prerequisites
• Installation acc. to IEC 60346-7-710: earth leakage circuit breaker (FI-circuit breaker) with rated leakage current <0.03 A
• Connection of the power supply cable of the ATMOS
max. 3 meters, preferably left (fig.). This may only be carried out by authorized qualified personnel.
• The supply circuit must be separated from other devices e.g. PCs etc.
• If isolating transformers are used then isolation monitoring must be integrated in the isolating transformer.
• Maximum power consumption:
-
ATMOS
S 61 Servant workstation 2.300 VA
®
- ATMOS
S 61 Servant vision 200 VA
®
- ATMOS
S 61 Servant instruments 250 VA
®
Flexible multi-port distributors may not be used as a power supply for the ATMOS@ S 61 Servant vision or ATMOS
Servant workstation.
6
160
Maximum required connections
1 x earthing contact socket outlet
1 x earthing contact socket outlet (basic version) or
1 x fixed connection for the integrated camera or stroboscope LED
3 x earthing contact socket outlet
1 x earthing contact socket outlet
1 x earthing contact socket outlet
1 x earthing contact socket outlet
1961
S 61 Servant
®
S 61 Servant vision to a safety socket outlet near the device,
®
1730
740
2940
S 61 Servant workstation, ATMOS
®
patient chair,
®
250
445
979,5
S
®
S 61
®