3.0 Setting up and starting up

Connection to central gas supply up to max. 8 bar.
The standard NIST connection for central vacuum and central compressed air on the rear side of the device is optional.
Connecting hoses are not included in delivery.

3.1 Required connections for all units of ATMOS S 61 Servant

Flexible multi-port distributors may not be used as a power supply for the ATMOS S 61 Servant ENT workstation, ATMOS S 61
Servant vision, or for an optional HF or radiofrequency surgical device.
Unit / Device
ATMOS S 61 Servant ENT workstation
ATMOS S 61 Servant vision
ATMOS S 61 Servant instruments
Water separation system
Monitor
Patient chair
An adequate number of socket outlets with earthing contact should be mounted for possible connection of further electrically
operated units that may be installed (installation of an electrically operated ATMOS patient chair, water separation system,
camera, monitor, etc.).

3.2 Connection to electrical power line

Prerequisites:
• Installation acc. to IEC 60346-7-710: earth leakage circuit breaker (FI circuit breaker) with rated leakage current < 0.03 A
• Connection of the power cable of the ATMOS S 61 Servant ENT workstation to a safety socket outlet near the device, max.
3 meters, preferably left (Fig.). This may only be carried out by authorized qualified personnel.
• The supply circuit must be separated from other devices, e.g., PCs, etc.
• If isolation transformers are used, then isolation monitoring must be integrated in the isolation transformer.
• Maximum power consumption:
- ATMOS S 61 Servant ENT workstation:
- ATMOS S 61 Servant vision:
- ATMOS S 61 Servant instruments:
• Flexible multi-port distributors may not be used as a power supply.
10
160
Maximum required connections
1 x earthing contact socket outlet
1 x earthing contact socket outlet (basic version) or
1 x fixed connection for the integrated camera or stroboscope LED
3 x earthing contact socket outlet
1 x earthing contact socket outlet
1 x earthing contact socket outlet
1 x earthing contact socket outlet
2,300 VA
300 VA
250 VA