5.0 Cleaning and care

5.1 General information on cleaning and disinfection

Prior to cleaning
Set main switch of the ATMOS
Prior to each use:
) The described action relating to cleaning and disinfection
resp. sterilisation do not substitute the relevant
instructions which must be adhered to prior to operation!
• For disinfection, you may use all surface and instrument
disinfectants listed on page 10.
• Please take care that all surfaces are dry wiped. Use a
single-use cloth to absorb any liquid.
• In case that any disinfectant is spilled, please take care
to dry wipe the surface immediately in order that no liquid
may penetrate gaps and edges.

5.1.1 Cleaning the unit surface

• The surfaces of the ATMOS
are resistant against all surface disinfectants listed in on
page 11.
• Wipe the unit surface with a cloth moistened with a
cleaning or disinfecting solution.
• You may also use disinfectant sprays or disinfectant
tissues for cleaning and disinfection.
ª Please note that the alcohol contained in these
agents could corrode or cloud the protective covers if
employed on a long term basis.
• Only deposit clean instruments on the board!
• Clean and disinfect the instruments regularly!
S 61 Servant instruments to OFF prior to cleaning and disinfection!
®
S 61 Servant instruments
®
if required
) Always observe the concentration specifications and
instructions by the respective manufacturer!
• Do not use
- Disinfectants which contain organic or inorganic acids
or bases as they could cause corrosion damage.
- Disinfectants containing chloramides or phenol
derivatives, since these may cause stress cracks in
the material used for the housing of the unit.

5.1.2 Instrument trays

• Before disinfection, thoroughly rinse the trays under
running water. A detergent or cleaning agent (surface
disinfectant) may also be used if required.
- Use water to thoroughly rinse all residues of these
substances.
• Melamine and anodized aluminium trays cannot be
sterilised.