Hygiene Plan - Atmos S61 Servant Operating Instructions Manual

Instruments
Hide thumbs Also See for S61 Servant:
Table of Contents

Advertisement

6.0 Hygiene Plan

What
Reusable parts
Secretion canister
Hose connection (nozzle)
Suction lid
Gasket
Bacterial fi lter
Splash protection
Float ball
Suction hose in the
canister
Secretion canister
Disposable canister
system
Hose irrigation system
Suction nozzle
for hose irrigation
Silicone attachment piece
Suction nipple
Secretion suction hose
Storage canister for
hose irrigation system
Ear irrigation / Thermal nystagmus stimulation
Ear irrigation bowl
Handle
Jet connection
Splash protection
Hose tip (disposable)
Rinsing attachment
Hygiene fi lter
Rinsing lid with rinsing
hose
Rinsing bottle
Medication nebulisation / Politzer
Handle for
compressed air
Sprayer jet
Sprayer head
Hose at sprayer head
Sprayer bottle
Politzer olive
Politzer connection
Endoscope management
Plastic quiver
Metal quiver
Fixation adapter for
plastic quiver
Protective sleeve (tefl on
element for metal quiver)
Cleaning and disinfection plan
ATMOS
How
C
D
S
Cleaning
Disinfection
Sterilisation
Cleaning and disinfection (automatic or manual)
X
X
2,4,5
Cleaning and disinfection (automatic or manual)
X
X
2,4,5
Cleaning and disinfection (automatic or manual)
X
X
2,4,5
Daily exchange or when fi lter is blocked
Cleaning and disinfection (automatic or manual)
X
X
2,4,5
Cleaning and disinfection (automatic or manual)
X
X
X
X
Cleaning and disinfection (automatic or manual)
2,4,5
Empty when the canister is full; at least daily
X
X
Cleaning and disinfection (automatic or manual)
Exchange and disposal of full canister
Wipe cleaning and disinfection
X
X
3
Cleaning and disinfection (automatic or manual)
X
X
2,4,5,6
Exchange of the Silicone attachment
X
Manual cleaning after each patient
X
Manual or automatic cleaning and disinfection
2,4,5,6
Rinse the secretion hose with the hose irrigation
X
system after each procedure;
Exchange or disinfection of the hose
X
2,4,5,6
Cleaning with a brush; Manual or automatic cleaning and
X
X
2,4,5,6
disinfection
Manual or automatic cleaning and disinfection
X
X
2,4,5
X
X
Wipe cleaning and wipe disinfection
3
Manual or automatic cleaning and disinfection
X
X
2,4,5,6
Cleaning and disinfection (automatic or manual)
X
X
2,4,5
Exchange after each application
Manual or automatic cleaning and disinfection
X
X
2,4,5
See operating instructions for hygiene fi lter
Manual or automatic cleaning and disinfection
X
X
2,4,5
Cleaning and disinfection; cleaning in the dishwasher
X
X
2,4,5,6
with the glass care programme
Manual cleaning and disinfection
X
X
3
X
Cleaning after each application
X
Manual or automatic cleaning and disinfection
2,4,5,6
Multiple rinsing of the sprayer head with water
X
2,4,5
Weekly exchange of the hose or when changing the
X
X
medication
Cleaning in a cleaning and disinfection device; weekly or
X
X
2,4,5,6
when changing the medication
Exchange after each application.
X
X
2,4,5,6
Cleaning and disinfection
Exchange after each application.
X
X
2,4,5,6
Cleaning and disinfection
Cleaning with a brush; disinfection
X
X
2,4,5
Cleaning with a brush; disinfection (automatic or manual)
X
X
2,4,5,6
X
X
Manual or automatic cleaning and disinfection
2,4,5
Manual or automatic cleaning and disinfection
X
X
2,4,5
S 61 Servant
®
Recommendations
When
Who
Qualifi ed and trained
staff who are familiar
with reprocessing.
(Please fi ll in the
responsible person
-> use a water-based
overhead marker)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
13