3.0 Setting up and starting up

3.1 Required connections (power supply)
Unit / Device
ATMOS
C 11 Systema
®
Optional:
Monitor
Microscope ATMOS
i View
®
Patient chair

3.2 Connection to electrical power line

• According to the directions of VDE 0107 and VDE 0100, medically used rooms have to be equipped with a leakage current
protective circuit (FI protective circuit) with a nominal leakage current of < 0.03 A. The installation must correspond with VDE
0107.
• The power cable of the ATMOS
unit (max. 5 m, see fig.). This may only be carried out by authorized qualified personnel.
• The power consumption comes to 270 W max.
• For the connection of further electrical devices, please allow for extra plugs (installation of an electrically operated patient
chair, etc.).
• In order to switch on the unit bring the on/off switch into "on" position.
• In order to switch off the unit bring the on/off switch into "off" position. In this position the unit is disconnected from the power
supply.
1100
960
1160
2260
C 11 Systema is connected to an earthing contact socket which is to be mounted near the
®
1519
800
1500
Maximum required connections
1x earthing contact socket outlet
1x earthing contact socket outlet
1x earthing contact socket outlet
1x earthing contact socket outlet
7