3 Setting up and starting up

3.1 Required connections (power supply)

Unit / Device
ATMOS C 11 Systema
Optional:
Monitor
Microscope ATMOS i View
Patient chair

3.2 Connection to electrical power line

• According to the directions of VDE 0107 and VDE 0100, medically used rooms have
to be equipped with a leakage current protective circuit (FI protective circuit) with
a nominal leakage current of < 0.03 A. The installation must correspond with VDE
0107.
• The power cable of the ATMOS C 11 Systema is connected to an earthing contact
socket near the device (max. 5 m; see figure). This may only be carried out by author-
ized qualified personnel.
• The maximum power consumption is 270 W.
• Please provide for additional sockets for the connection of further electrical devices
(installation of an electrically operated ATMOS patient chair, etc.).
• To switch on the unit, move the on/off switch to the "on" position.
• To switch off the unit, move the on/off switch to the "off" position. In this position,
the unit is disconnected from the power supply.
12 Setting up and starting up
Maximum required connections
1x earthing contact socket outlet
1x earthing contact socket outlet
1x earthing contact socket outlet
1x earthing contact socket outlet