2.0 For your safety

For your safety
• The ATMOS
S 61 Servant instruments is produced
®
according to IEC 601 / EN 60601 and listed in the
following classes:
- VDE Class of protection 1
- Class IIa (EEC 93/42).
• The device may only be connected to a properly installed
grounded electrical outlet.
Correct configuration in assembly of country-specific
connections:
green/yellow: non-fused earth conductor (PE)
blue: neutral conductor (N)
black resp. brown: phase (L)
• Caution! Mirror and endoscope heaters may generate
temperatures above 40°C!
• The ATMOS
S 61 Servant instruments may only be
®
used under the supervision of skilled staff who have been
authorised by ATMOS and trained in its operation (IEC
601-1 / EN 60601-1).
• The mains voltage indicated on the type plate must
correspond to the values of the supply network.
• Make sure prior to every application of the equipment that
it is technically safe and in proper condition. Damaged
leads and hoses must be replaced immediately!
• It is prohibited to simultaneously pull out several heavily
laden instrument deposits.
• Never leave the patient unattended at the treatment unit.
!
• Please note:
A medical insulating transformer with earth leakage
monitor or any similar safety system acc. to EN 60 601-1
is required, if several devices are connected over one
common power supply. The transformer must correspond
to the power consumption of all the devices to be
connected.
• Do not place used contaminated instruments on the ENT
unit except on destined places!
• The ambient conditions specified in the "Technical data"
must be strictly observed!
• Switch off main switches of the installed devices after
finishing work in the practice.
• The ATMOS
S 61 Servant instruments may be operated
®
only in rooms used for medical purposes, but not in
areas subject to explosion hazards and in oxygen rich
environments.
• The ATMOS
S 61 Servant instruments meets the
®
immunity to interference requirements of IEC 601-1-2 /
EN 60601-1-2 „Electromagnetic Compatibility – Medical
Electrical Devices".
• The ATMOS
S 61 Servant instruments may not
®
be operated with devices not complying with the
requirements of standard EN 60601-1 „Medical Electrical
Equipment" and EN 60601-1-2 „Electromagnetic
Compatibility" (Medical Electrical Equipment).
• ATMOS is not liable for personal injury and damage to
property if
- no original ATMOS parts are being used,
- the advice for use in these operating instructions is
not being observed,
- assembly, new settings, alterations, extensions
and repairs have been carried out by personnel not
authorised by ATMOS.
• In order to fully disconnect the device from the mains
supply the mains plug must be removed from the wall
outlet.
• This product is not re-sterilizable. Repeated reuse of
components which are marked with a
In case of repeated reuse these components lose their
function and there is a high infection risk.
2
is forbidden.