3.0 Setting up and starting up
Fig. 5.

3.3 Connections

3.3.1 Connection to electrical power line
• According to the directions of VDE 0107 and VDE 0100,
medically used rooms have to be equipped with a leak-
age current protective circuit (FI protective circuit) with
a nominal leakage current of < 0.03 A. Installation must
correspond with VDE 0107.
• Connect the power cable to the mains supply (, Fig. 2).
• Insert the power plug in a correctly installed socket with
earthing contact.
3.3.2 Connecting a equipotential bonding connec-
tion
• Connection for potential equalization (, Fig. 2). Con-
necting cable available from ATMOS (see section 8.0).

3.3.3 Water connection

• Local requirements:
- Water tap with G3/4" external thread.
- Drinking water!
- Input water pressure: +2000...+6000 hPa (6 bar).
- Feed temperature: +7...+19.5 °C (recommended),
but at least 0.5 °C below the required lowest cold
stimulation value.
- Drainage fittings (art. no. 502.0880.0)
• Connection to water supply and waste water:
- There are country-specific regulations for the instal-
lation to be considered when the unit is connected
to the public water supply. To comply with the rules
according to EN 1717, an unpressurized water sepa-
ration with overflow is required.
Therefore, ATMOS recommends installing the water
separation unit ATMOS Aqua Clean, REF 502.1200.0,
between the water supply and the device in order to
meet the connection requirements according to EN
1717.
- Prior to connecting the unit to the water supply, the
feed line must be flushed clean by opening the water
tap for a minute and allowing free flow of the tap
water.
- Connect the filter unit to the water tap by means of
adapter and gasket.
- Insert the gaskets in the cap nuts of the water supply
hose.
- Join the nuts with the filter connection and the unit
connection (, Fig. 3).
- Insert the gaskets in the cap nuts of the waste water
hose.
- Join the nuts with the unit connection (, Fig. 3) and
the drainage fittings.
- There is no special calcification safety device inte-
grated in the water system. Such a system is to be
connected when the respective drinking water is of
hardness grade 3 (14 - 21°d resp. 2.5 - 3.8 mmol/l
= hard water) and with hardness grade 4 (from 21°d
resp. from 3.8 mmol/l = very hard water).
Please contact your local water supplier and/or
plumber.
Water
Millimol per
hardness
litre
1 (soft)
≤ 1.3
2 (medi-
1.3 to 2.5
um)
3 (hard)
2.5 to 3.8
4 (very
> 3.8
hard)
!
Close the water tap when the unit is not in use!
°dH
≤ 7.3
Calcification
protection system
7.3 to 14
is not required
14 to 21.3 Calcification
protection system
> 21.3
is required