Single Lead vs. Multi-Lead:
This option examines ECG
leads utilized for arrhythmia
analysis.
•
Multi-Lead Analysis examines
ECG leads I, II, III and V to
help eliminate false alarms
and improve the ability of the
system to:
–
Detect beats that occur
isoelectric to a single chest
lead.
Discriminate artifact that
–
appears in one lead
compared to other lead
vectors.
–
Provides a "Smart Lead
Fail" feature where the
failed lead is identified and,
if available, another lead is
provided for display.
–
Continue arrhythmia
processing after a lead
change.
•
Single Lead Analysis uses
only the top displayed lead
to process heart rate and
arrhythmia information.
•
Single lead analysis can be
beneficial when
troubleshooting heart rate
and arrhythmia alarms.
•
To Change to Single Lead
Analysis:
Select the ECG Parameter
–
Window.
Select LD ANALYSIS:
–
MULTI-LEAD.
–
Move the cursor in front of
Single Lead and press to
select.
–
Select MAIN MENU to
exit.
TAB
TRIM
COVER &
NON-TAB
TRIM
COVER &
NON-TAB
TRIM
TAB
TRIM