GE DASH 3000 V5 Quick Reference Manual page 99

Bedside monitoring
Table of Contents

Advertisement

Problem:
Why does the monitor continue to read INFLATE
BALLOON after the balloon has been inflated?
Answer:
The monitor must detect a 30% change in the waveform
to measure a wedge. If the waveform does not change
accordingly, the WEDGE PROCESSING message will
not appear.
Intervention:
It would be recommended to complete a PA Wedge
utilizing the manual mode.
Invasive Pressures
Problem:
Why is the invasive waveform overshooting the scales?
Answer:
A new scale needs to be selected to accomodate the
change of the pressure reading. If Auto Scale was
previously selected, this would have used the current
waveform reading as a reference when this function was
requested. If the waveform readings have fluctuated, a
larger range scale or reselecting Auto Scale will reset the
scale for the current reading.
Problem:
When using a UAC, does it make a difference to label it
as an ART instead of UAC?
Answer:
No, it does not. But remember when looking up the
information under VITAL SIGNS or GRAPHIC
TRENDS, no data will be found under the displayed
label.
TROUBLESHOOTING and
FREQUENTLY ASKED QUESTIONS
TAB
TRIM
COVER &
NON-TAB
TRIM
COVER &
NON-TAB
TRIM
TAB
TRIM

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dash 4000 v5

Table of Contents