Modules
CO
: End-tidal CO
monitoring
2
2
is a continuous, non-invasive
technique for determining the
concentration of carbon dioxide.
The CO
level is measured by
2
the absorption of infrared light
of specific wavelengths in
respiratory gases.
Mainstream ETCO
Setup:
2
•
Securely insert module into
the Dash monitor.
•
Connect capnostat sensor
cable to module.
NOTE: Once the CO2 cable is connected to the module, a
"Warming Up" message is displayed on the CO2 parameter
window, which takes approximately two minutes.
To CAL SENSOR TO ZERO
•
CELL:
–
Remove adapter from
sensor.
–
Ensure cell windows are
clean and dry.
–
Place sensor on the cell
marked "0".
Select the CO
parameter
–
2
window.
–
Select the CAL SENSOR
TO ZERO CELL option
from the CO
parameter
2
window.
Popup menu opens with
–
READY or ABORT.
Select READY. The
–
message "Calibrating"
appears. When calibration
is complete, "0" appears.
–
Remove sensor from the
"0" cell and place it on the
REF cell. An EXP value of
38 mmHg (+/– 2 mmHg) is
displayed. If the value is
not within this range, the
sensor is out of tolerance
and should be replaced.
•
Snap mainstream adapter into
sensor and attach to ventilator
system.
Adapter calibration is required if:
•
Adapter types are changed.
•
Monitor displays: "CHECK
ADAPTER/ADAPTER CAL"
To calibrate the adapter:
•
–
Place the sensor and
adapter away from all
sources of CO
.
2
Optional Equipment
With the adapter connected
–
to the sensor, and the
sensor connected to the
monitored, select the
CALIBRATE ADAPTER
option from the CO
menu.
2
A popup menu opens with
READY or ABORT.
–
Select READY. The
message CALIBRATING is
displayed in the CO
2
parameter window. When
the calibration is complete,
the popup menu closes and
the message clears.