SPO
— Masimo SET allows the
2
sensitivity and averaging times
to be adjusted:
•
Sensitivity options include:
–
Normal, for routine patient
monitoring purposes.
–
Maximum for improved low
perfusion performance and
for faster tracking of SPO
changes.
–
Averaging menu options
determine how many
seconds the collected SPO
information is averaged.
Options for averaging times
are: 2, 4, 8, 10, 12, 14, or
16 seconds.
SPO
Menu — Masimo SET Probes
2
Rejuvenating the Adhesive
Properties of the Disposable
Saturation Probe:
2
•
Rejuvenate the probe at least
once per shift and PRN.
•
Rub the adhesive sides of the
probes with an alcohol
2
preparation pad.
•
After the probe has dried for
at least one minute, replace it
on the site.
NOTE: It is recommended that the adhesive tapes be changed
with every site change. Adhesive tape can not be rejuvenated.