Invasive Pressures —
Smart BP:
Smart BP is a feature found in
the ART and FEM pressure
menus that reduces nuisance
alarms associated with zeroing
the transducer, fast flushing
and drawing blood.
To Turn Smart BP On or Off:
Select ART or FEM
parameter window.
Select SMART BP and
press to select.
On/Off status is noted in
the menu option.
Select MAIN MENU to
exit.

Disconnect Alarm:

A feature found in the ART
and FEM pressure menus.
If the mean pressure falls
below 25 mmHg and the
disconnect alarm is on, a
Warning Alarm sounds and
the message
DISCONNECTED is
displayed in the parameter
window.
Check the patient
immediately in the event the
catheter has dislodged.
Smart BP
Disconnect Alarms