PA
PA Insert/Wedge is a feature
found in the PA Invasive
Pressures Menu. There are two
modes for doing a PA Wedge
measurement: Auto and
Manual.
The Auto Mode allows a PA
Wedge measurement to be
performed without having to
touch the monitor after
starting the program. The
Manual Mode defaults when
the PA Insert/Wedge menu
option is selected.
NOTE: The monitor must detect a 30% change in the
waveform to measure a wedge. If the waveform does not
change accordingly, the Wedge Processing message will not
appear and you must use the manual mode for wedge
measurements.
•
To Complete an Auto PA
Wedge:
–
Select PA.
Select PA INSERT/
–
WEDGE.
–
Select MODE:MANUAL
(this changes the mode to
Auto).
–
An INFLATE BALLOON
message appears in the PA
parameter window.
–
Deflate the balloon.
PA Wedge Review
–
Information window is
automatically displayed.
PA Wedge Review
–
To change the PA value,
select MOVE WEDGE
CURSOR.
–
Turn the Trim Knob to
move the cursor to the
desired position and press
to select.
PA Insert/Wedge
–
Press the GRAPH GO/
STOP button on the
monitor to print window.
–
Select MAIN MENU to
exit.
NOTE: The last PAW with a time stamp will be displayed in the
PA parameter window.