Invasive Pressures — IABP:
The IABP program
compensates for the
irregularities in the pressure
waveform caused by the use of
an intra-aortic balloon pump.
To ensure that the trigger
signal is compatible with all
modes of the IABP, the signal
source used to trigger an
intra-aortic balloon pump
should be the pump itself.
If the trigger is from the
monitor, be certain that the
manufacturer is compatible
with the GE Medical Systems
Information Technologies analog
output signal.
If the trigger is off of the
R-Wave, review the patient's
ECG leads and place the one
with the greatest amplitude in
the display (top) lead
position.
If blood pressure is used to
trigger the balloon, the first
red pressure port labeled
ART will be used.
Cable the balloon pump to
the monitor through the
Defib Sync connector on the
back of the monitor.
The Parameter Window
Displays:
Systolic Value = Highest
pressure in one cardiac
cycle.
Diastolic Value = Lowest
pressure in one cardiac
cycle.
Mean Value = Average
pressure in one cardiac
cycle.
NOTE: The values displayed will differ depending on the
timing of the pump.
To Turn the IABP Program
On:
Select ART 1.
Select IABP.
On/Off status is noted in
the menu option.
Select MAIN MENU to
exit.