GE DASH 3000 V5 Quick Reference Manual page 83

Bedside monitoring
Table of Contents

Advertisement

Troubleshooting
(continued)
PROBLEM
The monitor IS NOT
The monitor does not support the
communicating with the
auxiliary device.
auxiliary device.
The auxiliary cable is damaged.
The auxiliary device may need to be
configured to function with the
docking station.
The AC power indicator on
The docking station is not connected
the monitor and docking
to an electrical power outlet.
station IS NOT illuminated.
The electrical power outlet does not
have power.
The electrical power cable to the
docking station is defective.
The docking station is defective.
CAUSE
Use a monitor that supports the auxiliary device.
1. Disconnect the auxiliary cable from the docking station.
2. Connect the auxiliary cable to the monitor.
3. If the monitor cannot communicate with the auxiliary device, replace the
auxiliary cable.
See the Dash Patient Monitor operator' s manual and the operator's manual
provided with the auxiliary device for configuration recommendations.
Connect the docking station to an electrical power outlet.
1. Verify the electrical power outlet is functioning by connecting a known
good device to the outlet.
2. If the known good device does not work, then plug the docking station
into a known good electrical power outlet.
1. Verify the electrical power cable is defective by connecting a known
good electrical power cable to the docking station.
2. If the docking station's AC power indicator is now illuminated, replace
the defective power cable.
1. Verify the docking station is defective by connecting a known good
docking station.
2. If the docking station's AC power indicator is now illuminated, replace
the defective docking station.
SOLUTION
TAB
TRIM
COVER &
NON-TAB
TRIM
COVER &
NON-TAB
TRIM
TAB
TRIM